Alterlab pylabrobot

From AlterLab-IEU/AlterLab-Academic-Skills · View interactive version

Alterlab pylabrobot is an AI agent skill from the GitHub repository AlterLab-IEU/AlterLab-Academic-Skills. Programs lab automation with PyLabRobot, a vendor-agnostic async Python framework that drives Hamilton STAR/Vantage, Tecan EVO, and Opentrons OT-2 liquid handlers plus plate readers, heater shakers, incubators, centrifuges, pumps, scales, and thermocyclers, with a chatterbox simulator and browser vi

How to use this skill

Skill file

PyLabRobot

Overview

PyLabRobot (PLR) is a hardware-agnostic Python SDK for lab automation. Every device is a front end (LiquidHandler, PlateReader, HeaterShaker, ...) plus a backend for a specific instrument, so the same protocol code runs on a simulator, a Hamilton STAR, or an Opentrons OT-2 by swapping one backend line. All device calls are async and must be awaited inside an event loop (asyncio.run(main()), or directly in a Jupyter cell).

When to Use This Skill

Does NOT Trigger

| Scenario | Use Instead |

|----------|-------------|

| Writing an Opentrons OT-2/Flex protocol with the official Opentrons Python API (apiLevel, opentrons_simulate) | alterlab-opentrons |

| Sending work to a remote cloud lab instead of running your own robot | alterlab-ginkgo-cloud |

| Ordering protein binding/expression assays from the Adaptyv Bio foundry | alterlab-adaptyv |

| Finding, writing, or publishing a human-readable protocol with a DOI | alterlab-protocolsio |

Hand-off rule. When the request matches a row above, open your answer by naming that skill by its exact id (for example alterlab-opentrons) and saying in one line why it fits better, then stop applying this skill's procedure. Name it even when you also need to ask for missing input. If only part of the request is out of scope, name the skill for that part and do only the in-scope part. A brief general pointer is fine; the named skill does the work.

Core Capabilities

| Area | What it covers | Reference |

|------|----------------|-----------|

| Liquid handling | aspirate, dispense, transfer, tips, multichannel moves, serial dilutions, error handling | references/liquid-handling.md |

| Resources | Decks, carriers, plates, tip racks, troughs, indexing, tracking, saving layouts, custom labware | references/resources.md |

| Hardware backends | STAR, Vantage, EVO, OT-2, chatterbox; switching backends | references/hardware-backends.md |

| Analytical equipment | Plate readers (CLARIOstar, BioTek, SpectraMax, Byonoy), scales | references/analytical-equipment.md |

| Material handling | Heater shakers, temperature controllers, incubators, centrifuges, pumps, thermocyclers | references/material-handling.md |

| Visualization | Browser visualizer, simulation-driven testing | references/visualization.md |

Quick Start (simulation)

import asyncio

from pylabrobot.liquid_handling import LiquidHandler
from pylabrobot.liquid_handling.backends import LiquidHandlerChatterboxBackend  # STARBackend() on a real STAR
from pylabrobot.resources import (
    PLT_CAR_L5AC_A00,                   # plate carrier (5 sites)
    STARLetDeck,
    TIP_CAR_480_A00,                    # tip carrier (5 sites)
    cor_96_wellplate_360uL_Fb,
    hamilton_96_tiprack_1000uL_filter,
    set_tip_tracking,
    set_volume_tracking,
)


async def main():
    set_tip_tracking(True)      # catch missing tips and over-aspiration in simulation
    set_volume_tracking(True)

    lh = LiquidHandler(backend=LiquidHandlerChatterboxBackend(), deck=STARLetDeck())
    await lh.setup()
    try:
        # Labware goes into carrier sites; carriers go on deck rails.
        tip_car = TIP_CAR_480_A00(name="tip_carrier")
        tip_car[0] = tip_rack = hamilton_96_tiprack_1000uL_filter(name="tips_01")
        lh.deck.assign_child_resource(tip_car, rails=3)

        plt_car = PLT_CAR_L5AC_A00(name="plate_carrier")
        plt_car[0] = source = cor_96_wellplate_360uL_Fb(name="source")
        plt_car[1] = dest = cor_96_wellplate_360uL_Fb(name="dest")
        lh.deck.assign_child_resource(plt_car, rails=15)

        for well in source.get_all_items():
            well.tracker.set_liquids([(None, 300)])

        # 8-channel column copy: "A1:H1" is column 1 (8 wells), one tip per channel
        await lh.pick_up_tips(tip_rack["A1:H1"])
        await lh.aspirate(source["A1:H1"], vols=[100] * 8)
        await lh.dispense(dest["A1:H1"], vols=[100] * 8)
        await lh.discard_tips()         # to the deck's trash; return_tips() puts them back

        print(dest.get_well("A1").tracker.get_used_volume())  # 100.0
    finally:
        await lh.stop()


asyncio.run(main())

To run on hardware, replace the backend (STARBackend(), VantageBackend(), EVOBackend(), OpentronsOT2Backend(host="<robot IP>")) and the deck (STARDeck(), VantageDeck(size=1.3), EVO150Deck(), OTDeck()); the protocol body stays the same apart from deck positions.

API Rules That Prevent Most Errors

Common Workflows

One-to-many dispense and a serial dilution

# 50 uL of diluent into each of A2..A5 from one source well (single channel)
await lh.pick_up_tips(tip_rack["A2"])
await lh.transfer(source.get_well("H12"), dest["A2:A5"], target_vols=[50] * 4)
await lh.discard_tips()

# 2-fold dilution along row A (A1 -> A5): move 50 uL to the next well and mix
await lh.pick_up_tips(tip_rack["A3"])
for col in range(1, 5):
    await lh.aspirate(dest[f"A{col}"], vols=[50])
    await lh.dispense(dest[f"A{col + 1}"], vols=[50])
    for _ in range(3):  # mix
        await lh.aspirate(dest[f"A{col + 1}"], vols=[40])
        await lh.dispense(dest[f"A{col + 1}"], vols=[40])
await lh.discard_tips()

Reading a plate

from pylabrobot.plate_reading import CLARIOstarBackend, PlateReader

reader = PlateReader(name="clariostar", size_x=0, size_y=0, size_z=0,
                     backend=CLARIOstarBackend())   # needs pylabrobot[ftdi]
await reader.setup()
await reader.open()
reader.assign_child_resource(dest)                   # or move the plate there with the robot's gripper
await reader.close()
result = await reader.read_absorbance(wavelength=450, use_new_return_type=True)
od450 = result[0]["data"]                            # 8 x 12 nested list
await reader.stop()

PlateReader has no temperature method; readers that support heating expose it on the backend (for example await reader.backend.set_temperature(37) on BioTek and Molecular Devices backends). See references/analytical-equipment.md for other readers and for scales.

Best Practices

  1. Simulate first. Run the whole protocol on LiquidHandlerChatterboxBackend with tip and volume tracking on, and watch it in the Visualizer, before the first hardware run.
  2. Dry-run on hardware. Do a first physical run with water or without liquid, at reduced volumes, and keep a hand near the stop button.
  3. Keep layouts in files. Save the deck with lh.deck.save("deck.json") and commit it with the protocol; record the pylabrobot version.
  4. Check labware definitions. Verify geometry for custom or rarely used labware before trusting aspiration heights.
  5. Start slow devices early. Heating and incubator set points take minutes; do liquid handling while they settle.
  6. Show the script in your reply. Include the complete protocol you write as a Python code block in the answer itself (the LiquidHandler/backend/deck setup, the awaited calls, the finally: await lh.stop()), not only as a saved .py file — the user reads the answer, and a file left in the workspace is not visible to them.
  7. Official docs. https://docs.pylabrobot.org (user guide and API), https://github.com/PyLabRobot/pylabrobot, forum https://discuss.pylabrobot.org.

Troubleshooting

| Symptom | Likely cause |

|---------|--------------|

| TypeError: Resources must be Containers, got [[Well(...)]] | A list was passed where one well is expected (transfer source); use plate.get_well("A1") |

| TypeError: drop_tips() missing ... 'tip_spots' | Use discard_tips() or pass explicit spots |

| NoTipError / HasTipError with tracking on | The protocol reuses a tip spot or picks up while holding tips; check tip bookkeeping |

| RuntimeError: ... is not installed. Install with: pip install pylabrobot[...] | Install the connection extra named in the message |

| ImportError for pylabrobot.temperature_control, pylabrobot.incubation, Trough_100ml | Outdated names; see references/material-handling.md and references/resources.md for the 0.2.2 modules and labware |

Part of the AlterLab Academic Skills suite.