780g pump adjustment

From dromlakhani/MD2SKILL · View interactive version

780g pump adjustment is an AI agent skill from the GitHub repository dromlakhani/MD2SKILL. Guide pump setting adjustments for the MiniMed 780G system based on CareLink data, using Medtronic's 3-step follow-up workflow. Use when a clinician has CareLink data for a 780G patient and wants to know what to change — which setting, by how much, and how to program it. Triggers include: \"what do

How to use this skill

Skill file

MiniMed 780G Pump Adjustment Guide (CareLink-Based)

BEFORE YOU BEGIN

The MiniMed 780G is a hybrid closed-loop system. SmartGuard Auto Mode adjusts basal insulin automatically and delivers Auto Correction boluses, but it cannot compensate for incorrect Bolus Wizard settings. Most therapy problems in Auto Mode stem from meal bolus gaps, not basal errors.

Prerequisites for reliable adjustment decisions:

Golden rule: Change ONE setting at a time. Review in 2 weeks.


THE 3-STEP FOLLOW-UP FRAMEWORK

| Step | Question | Report to Use |

|---|---|---|

| 1. Review therapy goals | Are TIR and Statistics targets met? | Assessment & Progress Report |

| 2. Identify the problem | What pattern is causing out-of-range glucose? | Meal Bolus Wizard + Weekly/Daily Review |

| 3. Adjust the setting | Which setting needs changing, and by how much? | Device Settings Report |


STEP 1 — REVIEW THERAPY GOALS (Assessment & Progress Report)

Time in Range targets (Battelino 2019):

| Metric | Goal | Flag |

|---|---|---|

| Time above 250 mg/dL | <5% | >5% = significant hyperglycaemia |

| Time above 180 mg/dL | <25% | >25% = hyperglycaemia problem |

| TIR 70–180 mg/dL | >70% | <70% = action needed |

| Time below 70 mg/dL | <4% | >4% = hypoglycaemia problem |

| Time below 54 mg/dL | <1% | Any value = urgent |

Statistics targets:

| Statistic | Goal | If Not Met |

|---|---|---|

| SmartGuard use | ≥85% | Identify and fix SmartGuard Exits |

| Sensor wear | ≥85% | Counsel on CGM wear |

| GMI (Glucose Management Indicator) | Adults <7.0% / Paeds 7.0–7.5% | Surrogate A1C — flag if discordant with lab |

| CV (Coefficient of Variation) | <36% | ≥36% = unstable glucose, higher hypo risk |

| Auto Correction % of total bolus | <20–25% | >35% = meal bolus gap → adjust ICR |

| Set change interval | ≤3 days | >3.5 days avg = infusion site degradation |

Decision after Step 1:


STEP 2 — IDENTIFY THE PATTERN (Meal Bolus Wizard + Weekly/Daily Review)

2A. Determine the primary problem

Is it mainly HYPER, mainly HYPO, or both?

Look at the TIR bar on the Assessment & Progress report:

2B. Identify WHEN the glucose goes out of range

Use the Percentile Comparison graph (Section 4 of A&P Report):

Use the Weekly Review / Daily Review:

Use the Meal Bolus Wizard Report:

2C. Check SmartGuard Exits (Section 5 of A&P Report)

Preventable exits (address with patient education):

| Exit Reason | Cause | Action |

|---|---|---|

| BG required for SmartGuard | Patient not entering BG when prompted | Counsel: enter BG immediately when pump alerts |

| No SG values | Sensor gap — worn incorrectly or disconnected | Counsel on consistent sensor placement |

| Sensor Expired | Sensor left in >7 days | Counsel: change sensor on day 6–7 |

| SmartGuard disabled by user | Patient manually exiting Auto Mode | Discuss why; address barriers |

| Prolonged Suspend | Manual suspend >2h | Counsel: resume promptly after suspension |

Frequent preventable exits = SmartGuard use <85% = system cannot achieve optimal control. Address exits before adjusting settings.


STEP 3 — ADJUST THE SETTING

PRINCIPLE: ONE CHANGE AT A TIME. Never adjust two settings simultaneously.


3A. HYPERGLYCAEMIA — Setting Adjustments

Problem: High Auto Correction % (>35%) + Postprandial Highs

Diagnosis: Insulin-to-Carb Ratio (ICR) is too low — not enough insulin for meals

Setting to adjust: Carb Ratio (g/U)

How to program on the pump:

Home screen → Press ⊙ → Settings ⚙ → Delivery Settings → Bolus Wizard Setup → Carb Ratio

Select the relevant time period, enter new g/U value, select Done → Save


Problem: Consistent Fasting/Overnight Hyperglycaemia (not postprandial)

Diagnosis: Dawn phenomenon OR overnight SmartGuard target too low for this patient

Option A — Raise SmartGuard glucose target (first-line in Auto Mode)

How to program on the pump:

Home screen → Press ⊙ → Settings ⚙ → SmartGuard → Target

Select new target → Save

Option B — Increase overnight basal rate (if issues persist in Manual Mode or SmartGuard exits frequent overnight)

How to program on the pump:

Home screen → Press ⊙ → Settings ⚙ → Delivery Settings → Basal Pattern Setup

Select active pattern → Options → Edit → Navigate to the 00:00 segment → Adjust U/hr → Done → Save


Problem: Consistent Post-Correction Highs (correction boluses not working well)

Diagnosis: Insulin Sensitivity Factor (ISF) too low — corrections are not strong enough

Setting to adjust: Insulin Sensitivity Factor (ISF / Sensitivity)

How to program on the pump:

Home screen → Press ⊙ → Settings ⚙ → Delivery Settings → Bolus Wizard Setup → Sensitivity (ISF)

Enter new mmol/L (or mg/dL) per U value → Done → Save


Problem: Glucose Rises 3–4 Days After Set Change

Diagnosis: Infusion site degradation — not an insulin dose problem

Action: Counsel on ≤3-day set rotation. Set a reminder on the pump.

Settings → Reminders → Set Change → Set to 3 days

No insulin dose change needed.


3B. HYPOGLYCAEMIA — Setting Adjustments

⚠️ Address hypoglycaemia before any hyperglycaemia changes. Safety first.

Problem: Post-Meal Lows (<3h after bolus)

Diagnosis: ICR too aggressive (too much insulin for food) OR bolus given too early

Setting to adjust: Carb Ratio (g/U)

How to program: Same path as above → Delivery Settings → Bolus Wizard Setup → Carb Ratio


Problem: Fasting/Between-Meal Lows

Diagnosis: Basal rate too high

Setting to adjust: Basal Pattern

How to program: Same path as basal adjustment above → Basal Pattern Setup


Problem: Frequent Low Glucose Alerts + High CV

Diagnosis: ISF too aggressive; or Active Insulin Time too short (insulin stacking)

Option A — Increase ISF (less aggressive correction)

Option B — Extend Active Insulin Time (AIT)

How to program AIT:

Home screen → Press ⊙ → Settings ⚙ → Delivery Settings → Bolus Wizard Setup → Active Insulin Time

Set duration → Save


Problem: Nocturnal Lows (11PM–5AM)

Diagnosis: Overnight basal too high OR evening correction bolus stacking

Adjustment: Reduce overnight basal rate by 10–20%. Also counsel on evening bolus timing — avoid corrections within 2h of bedtime if AIT is 2 hours.


3C. HIGH GLYCAEMIC VARIABILITY (CV ≥36%)

High CV = unstable glucose = increased hypoglycaemia risk regardless of mean glucose.

Common causes:

Actions:

  1. Reinforce pre-meal bolusing (enter carbs + give bolus BEFORE eating)
  2. Consider dual wave bolus for high-fat/high-protein meals (only in Manual mode)
  3. Reduce SmartGuard exits (see Step 2C)
  4. Counsel on treating lows with precise carb amount (15g rule)
  5. Ensure AIT is correctly set for the insulin used (typically 2 hours for aspart/lispro)

QUICK REFERENCE — FINDING → ADJUSTMENT → PUMP PATH

| CareLink Finding | Problem | Setting to Change | Direction | Pump Path |

|---|---|---|---|---|

| Auto Correction >35% + post-meal highs | ICR too low | Carb Ratio ↓ g/U | Decrease g/U | Settings → Delivery → Bolus Wizard Setup → Carb Ratio |

| Fasting/overnight highs | Dawn phenomenon | SmartGuard Target ↑ | 100 → 120 mg/dL | Settings → SmartGuard → Target |

| Fasting/overnight highs (Manual mode) | Overnight basal low | Basal rate ↑ | +10–20% | Settings → Delivery → Basal Pattern Setup |

| Corrections undershoot target | ISF too low | ISF ↓ mg/dL per U | Decrease mg/dL/U | Settings → Delivery → Bolus Wizard Setup → Sensitivity |

| Post-meal lows | ICR too high | Carb Ratio ↑ g/U | Increase g/U | Settings → Delivery → Bolus Wizard Setup → Carb Ratio |

| Between-meal lows | Basal too high | Basal rate ↓ | −10–20% | Settings → Delivery → Basal Pattern Setup |

| Correction bolus stacking / lows | AIT too short | Active Insulin Time ↑ | +30 min | Settings → Delivery → Bolus Wizard Setup → AIT |

| Site change-related highs | Infusion site | Site rotation | ≤3 days | Settings → Reminders → Set Change |

| SmartGuard use <85% | Preventable exits | Educate patient | See exit type | — |

| CV ≥36% | Variability | Bolus timing | Pre-meal bolusing | — |


SETTING CHANGE LIMITS AND SAFETY GUARDRAILS

> Settings ⚙ → Delivery Settings → Max Basal/Bolus → Max Bolus

> Settings ⚙ → Delivery Settings → Max Basal/Bolus → Max Basal

Key guardrails:


OUTPUT FORMAT — ADJUSTMENT PLAN

Present adjustments as:

Patient / Date range:

WHAT THE DATA SHOWS:

PROPOSED SINGLE ADJUSTMENT:

PATIENT EDUCATION POINT:

REVIEW IN: [2 weeks — standard after any setting change]


SOURCE

Use this skill with your AI agent: Open on Skill Issue to copy, download, or add to your library. Works with Claude, ChatGPT, Gemini, Cursor, Windsurf and more.