Binding characterization is an AI agent skill from the GitHub repository FridrichMethod/awesome-skills. Guidance for SPR and BLI binding characterization experiments. Use when: (1) Planning binding kinetics experiments, (2) Troubleshooting poor/no binding signal, (3) Interpreting kinetic data artifacts, (4) Choosing between SPR vs BLI platforms.

How to use this skill

Skill file

Binding Characterization: SPR and BLI

SPR vs BLI Decision Matrix

| Factor | Choose SPR | Choose BLI |

|--------|------------|------------|

| Sensitivity | Small molecules, fragments (<500 Da) | Large complexes, antibodies |

| Throughput | Low-medium (serial) | High (96-well parallel) |

| Sample purity | Required (clogs fluidics) | Tolerates crude lysates |

| Kinetic resolution | Higher (better for fast kinetics) | Lower |

| Mass transport | More sensitive (may distort kon) | Less sensitive |

| Maintenance | High (fluidics system) | Low (dip-and-read) |

| Sample consumption | Higher (continuous flow) | Lower |

| Cost per experiment | Lower chip cost, higher run cost | Higher tip cost, lower run cost |

Key differences

SPR (Surface Plasmon Resonance)

BLI (Biolayer Interferometry)


Troubleshooting: Why BLI works but SPR doesn't

| Cause | Mechanism | Solution |

|-------|-----------|----------|

| Hydrophobic CDRs | Adsorb to SPR gold/dextran surface | Add 0.05% Tween-20, use CM7 chip with longer dextran |

| Aggregation | Mass transport artifacts in SPR fluidics | Filter sample (0.22μm), reduce ligand density |

| High instability | Degrades during continuous flow | Shorter cycle time, add stabilizers (trehalose 5%) |

| Charge mismatch | Nonspecific binding to charged dextran | Adjust buffer pH ±1 from pI, add BSA 1mg/mL |

| Slow dissociation | Long regeneration needed (damages ligand) | Use BLI (disposable tips) |

Why SPR works but BLI doesn't

| Cause | Mechanism | Solution |

|-------|-----------|----------|

| Small analyte | BLI less sensitive for <10 kDa | Use SPR with appropriate chip |

| Weak affinity (KD >10μM) | Fast dissociation in BLI dip | Increase analyte concentration |

| Low expression | Not enough signal | Increase biosensor loading |


Mass transport considerations

Mass transport limitation occurs when analyte cannot diffuse to the surface fast enough to maintain equilibrium. This distorts kinetic parameters.

Symptoms

When mass transport matters

Mitigation strategies

| Strategy | SPR | BLI |

|----------|-----|-----|

| Reduce ligand density | <200 RU for high-affinity | <0.5 nm shift loading |

| Increase flow rate | 50-100 μL/min | Increase shake speed (1000 rpm) |

| Use oriented immobilization | His-tag capture | Biotinylated ligand |

| Include in fitting | Mass transport model (kt) | Usually less critical |


Nonspecific binding mitigation

Buffer additives (ranked by effectiveness)

| Additive | Concentration | Mechanism | Best For |

|----------|---------------|-----------|----------|

| BSA | 0.5-1 mg/mL | Blocks hydrophobic sites | General use |

| Tween-20 | 0.02-0.05% | Prevents surface adsorption | Hydrophobic analytes |

| Trehalose | 1-5% | Stabilizes + blocks | Unstable proteins |

| Sucrose | 5% | BLI-specific blocker | BLI tips |

| Carboxymethyl dextran | 1 mg/mL | Competitive blocking | SPR with charged proteins |

| NaCl | 150-500 mM | Reduces ionic interactions | Charged proteins |

pH optimization

Reference subtraction

Always include:


Regeneration conditions

SPR regeneration scouting (try in order)

| Condition | Targets | Caution |

|-----------|---------|---------|

| 10 mM Glycine pH 2.0-2.5 | Most protein-protein | May denature ligand |

| 10 mM Glycine pH 1.5 | Strong interactions | Harsh, limit exposure |

| 1-2 M NaCl | Ionic interactions | Mild, try first |

| 10 mM NaOH | Very stable ligands | Can hydrolyze proteins |

| 10 mM Glycine pH 9-10 | Acid-stable proteins | Can aggregate |

| 10 mM EDTA | His-tag, metal-dependent | Strips Ni-NTA |

| 4 M MgCl2 | Hydrophobic interactions | Check ligand stability |

Regeneration protocol

  1. Start with mildest condition (high salt)
  2. Test 30s contact time
  3. Verify complete dissociation (return to baseline)
  4. Verify retained ligand activity (repeat binding)
  5. Use shortest effective contact time

BLI tips


Common artifacts and solutions

Biphasic binding

Symptoms: Two-rate association or dissociation

Causes:

Solutions:

Negative dissociation

Symptoms: Signal increases during dissociation phase

Causes:

Solutions:

Hook effect

Symptoms: Signal decreases at high analyte concentrations

Causes:

Solutions:


Kinetic data quality checklist

Before analysis

Fitting quality

Red flags


References

Platform comparisons

SPR protocols

Troubleshooting

Regeneration

Mass transport