Develop gc method

From pjt222/agent-almanac · View interactive version

Develop gc method is an AI agent skill from the GitHub repository pjt222/agent-almanac. Develop a gas chromatography method from scratch: define analytical objectives, select column chemistry, optimize temperature programming, choose carrier gas and detector, and validate initial system performance for target analytes in a given matrix.

How to use this skill

Skill file

Develop a GC Method

Systematic GC method dev: column pick + temp program + carrier + detector + initial perf check for volatile/semi-volatile analytes.

Use When

In

Required

Optional

Do

Step 1: Analytical Objectives

  1. List analytes + props (bp, polarity, MW).
  2. ID matrix + expected interferents/co-extractives.
  3. Specify LOD/LOQ, quant range, Rs for critical pairs.
  4. Method must meet regulatory (EPA 8260, USP, etc)?
  5. Doc throughput: max run time, inj vol, prep constraints.

→ Written spec: analytes + matrix + limits + Rs + regulatory/throughput.

If err: volatility data missing → estimate bp from structural analogs or scout run on mid-polarity col for elution order.

Step 2: Pick Column

Dimensions + phase via analyte polarity + separation diff.

| Column Type | Stationary Phase | Polarity | Typical Use Cases |

|---|---|---|---|

| DB-1 / HP-1 | 100% dimethylpolysiloxane | Non-polar | Hydrocarbons, solvents, general screening |

| DB-5 / HP-5 | 5% phenyl-methylpolysiloxane | Low polarity | Semi-volatiles, EPA 8270, drugs of abuse |

| DB-1701 | 14% cyanopropylphenyl | Mid polarity | Pesticides, herbicides |

| DB-WAX / HP-INNOWax | Polyethylene glycol | Polar | Alcohols, fatty acids, flavors, essential oils |

| DB-624 | 6% cyanopropylphenyl | Mid polarity | Volatile organics, EPA 624/8260 |

| DB-FFAP | Modified PEG (nitroterephthalic acid) | Highly polar | Organic acids, free fatty acids |

| DB-35 | 35% phenyl-methylpolysiloxane | Mid-low polarity | Polychlorinated biphenyls, confirmatory column |

  1. Analyte polarity ↔ phase: like dissolves like.
  2. Length (15-60 m): longer → more plates, longer runs.
  3. ID (0.25-0.53 mm): narrower → better eff, wider → more capacity.
  4. Film (0.25-5.0 um): thicker → retain volatiles longer.
  5. Complex matrices → guard col or retention gap.

→ Col spec (phase + L + ID + film) justified by analyte + Rs reqs.

If err: no single col resolves all → confirm col w/ orthogonal selectivity (DB-1 primary, DB-WAX confirm).

Step 3: Optimize Temp Program

  1. Initial oven ≤ bp of most volatile (hold 1-2 min → solvent focus).
  2. Linear ramp starts:
  1. Final temp 10-20 C above bp of least volatile.
  2. Final hold 2-5 min → full elution + bake-out.
  3. Co-eluting critical pairs → isothermal hold before elution, or slower ramp there.
  4. Verify total run time meets throughput.

→ Temp program (init + hold + ramp + final + hold) separates all targets in acceptable time.

If err: critical pairs still unresolved after ramp → revisit col (Step 2) or multi-ramp w/ slower rates in problem region.

Step 4: Pick Carrier Gas

| Property | Helium (He) | Hydrogen (H2) | Nitrogen (N2) |

|---|---|---|---|

| Optimal linear velocity | 20-40 cm/s | 30-60 cm/s | 10-20 cm/s |

| Efficiency at high flow | Good | Best (flat van Deemter) | Poor |

| Speed advantage | Baseline | 1.5-2x faster than He | Slowest |

| Safety | Inert | Flammable (needs leak detection) | Inert |

| Cost / availability | Expensive, supply concerns | Low cost, generator option | Very low cost |

| Detector compatibility | All detectors | Not with ECD; caution with some MS | All detectors |

  1. Default He for general + regulatory methods specifying He.
  2. H2 for faster or when He supply constrained; install H2 leak detection + interlocks.
  3. N2 only for simple separations or when cost primary.
  4. Flow → optimal linear velocity for gas + col ID.
  5. Measure actual velocity via unretained (e.g., methane on FID).

→ Carrier picked + flow at optimal velocity, verified by unretained peak.

If err: eff lower than expected → van Deemter curve (plate height vs velocity) over 5-7 flows for true optimum.

Step 5: Pick Detector

| Detector | Selectivity | Sensitivity (approx.) | Linear Range | Best For |

|---|---|---|---|---|

| FID | C-H bonds (universal organic) | Low pg C/s | 10^7 | Hydrocarbons, general organics, quantitation |

| TCD | Universal (all compounds) | Low ng | 10^5 | Permanent gases, bulk analysis |

| ECD | Electronegative groups (halogens, nitro) | Low fg (Cl compounds) | 10^4 | Pesticides, PCBs, halogenated solvents |

| NPD/FPD | N, P (NPD); S, P (FPD) | Low pg | 10^4-10^5 | Organophosphorus pesticides, sulfur compounds |

| MS (EI) | Structural identification | Low pg (scan), fg (SIM) | 10^5-10^6 | Unknowns, confirmation, trace analysis |

| MS/MS | Highest selectivity | fg range | 10^5 | Complex matrices, ultra-trace, forensic |

  1. Match detector to analyte chem + sensitivity.
  2. Quant in simple matrices → FID default (robust + linear + low maint).
  3. Trace in complex matrices → MS SIM or MS/MS MRM.
  4. Halogenated at trace → ECD best sensitivity.
  5. Detector temp 20-50 C above max oven → prevent condensation.
  6. Optimize detector gas flows per mfr.

→ Detector picked + config w/ temps + flows for targets.

If err: sensitivity insufficient → concentrate sample (bigger inj, solvent evap) or more sensitive/selective detector.

Step 6: Validate Initial Perf

  1. System suitability std: all targets at mid-range conc.
  2. Inject std 6× consec.
  3. Evaluate:
  1. Blank inj → no carryover/ghost peaks.
  2. Matrix blank → ID interferents at target RT.
  3. Doc all in method summary.

→ Suitability met across replicates, no carryover/matrix interference at target windows.

If err: tailing → check active sites (recondition, trim 0.5 m inlet, replace liner). RSD high → autosampler precision + inj technique. Rs low → Step 3 temp refinement.

Check

Traps

→

Use this skill with your AI agent: Open on Skill Issue to copy, download, or add to your library. Works with Claude, ChatGPT, Gemini, Cursor, Windsurf and more.