Interpret chromatogram is an AI agent skill from the GitHub repository pjt222/agent-almanac. Interpret a chromatogram from GC or HPLC analysis: verify system suitability parameters, identify peaks by retention time and spectral matching, perform accurate peak integration, calculate chromatographic figures of merit, and assess overall peak quality for reliable quantitation.

How to use this skill

Skill file

Interpret a Chromatogram

Systematic interpretation of GC and HPLC chromatograms covering system suitability verification, peak identification, integration, calculation of chromatographic parameters, and assessment of peak quality for confident qualitative and quantitative results.

When to Use

Inputs

Required

Optional

Procedure

Step 1: Verify System Suitability

Confirm that the chromatographic system is performing within specification before interpreting sample data.

| Parameter | Typical Specification | Calculation |

|---|---|---|

| Retention time RSD | <= 1.0% | RSD of tR over n >= 5 injections |

| Peak area RSD | <= 2.0% (assay), <= 5.0% (impurity) | RSD of area over n >= 5 injections |

| Tailing factor (T) | 0.8-2.0 (USP), ideally 0.9-1.2 | T = W0.05 / (2 * f) where W0.05 = width at 5% height, f = front half-width |

| Resolution (Rs) | >= 1.5 (baseline), >= 2.0 (regulated) | Rs = 2(tR2 - tR1) / (w1 + w2) |

| Theoretical plates (N) | Per column spec (e.g., >= 2000) | N = 16(tR / w)^2 or N = 5.54(tR / w0.5)^2 |

| Capacity factor (k') | 2.0-10.0 for primary analyte | k' = (tR - t0) / t0 |

  1. Locate the system suitability injections (typically 5-6 replicates of a reference standard at the start of the sequence).
  2. Calculate each parameter from the table above.
  3. Compare calculated values against the method's acceptance criteria.
  4. If any parameter fails, the system is not suitable -- do not proceed to sample interpretation until the issue is resolved.
  5. Document all system suitability results in the batch record.

Got: All system suitability parameters within specification, confirming the system is fit for purpose.

If fail: If retention time RSD fails, check for temperature instability, mobile phase preparation errors, or column degradation. If tailing factor fails, inspect the inlet liner (GC) or column frit (HPLC). If resolution fails, check column performance with a dedicated test mix and replace if necessary.

Step 2: Identify Peaks

  1. Compare each peak's retention time (tR) against the reference standard chromatogram.
  1. For ambiguous identifications, use co-injection (spiking): add reference standard to the sample and re-inject. The target peak should increase without broadening or shouldering.
  2. For DAD-equipped HPLC: compare the UV-Vis spectrum of each peak against a spectral library.
  1. For MS-equipped systems: confirm molecular ion (m/z) and key fragment ions against reference spectra.
  2. Flag any peak that cannot be identified -- report it as "unknown" with its retention time and relative response.

Got: All target analytes identified by retention time matching, with spectral confirmation where available. Unknown peaks flagged with retention time and area.

If fail: If retention times have shifted uniformly, a systematic change has occurred (column aging, temperature drift, mobile phase error). Re-inject the reference standard to establish current retention times before re-evaluating.

Step 3: Perform Peak Integration

  1. Select integration mode:
  1. Set integration parameters:
  1. Verify baseline placement:
  1. Check for integration errors:
  1. Record the final integration parameters and any manual adjustments with justification in an audit trail.

Got: All target peaks integrated with correct baseline placement, no artifacts included, and all manual adjustments documented with rationale.

If fail: If the automatic integrator consistently mishandles a particular peak shape, create a timed-events integration method with custom parameters for that retention window. Never manually adjust integration to achieve a desired result -- adjustments must be scientifically justified.

Step 4: Calculate Chromatographic Parameters

Calculate the following for all reported peaks:

  1. Resolution (Rs) between adjacent peaks:
  1. Tailing factor (T) at 5% peak height:
  1. Theoretical plates (N):
  1. Capacity factor (k'):
  1. Selectivity factor (alpha) between critical pair:
  1. Tabulate results for all analytes and compare against method specifications.

Got: All chromatographic parameters calculated, tabulated, and compared to acceptance criteria. Critical pair resolution and plate count documented.

If fail: If calculated plates are significantly below the column specification, the column may be degraded -- test with a fresh standard and compare to historical data. If parameters drift within a sequence, investigate instrument stability.

Step 5: Assess Peak Quality

  1. Symmetry: Peaks should be Gaussian or near-Gaussian. Significant fronting (T < 0.8) suggests column overload; tailing (T > 1.5) suggests secondary interactions or dead volume.
  2. Baseline separation: For quantitative work, critical pairs must be baseline-resolved. If valley between peaks does not return to baseline, note the percentage valley and assess impact on accuracy.
  3. Peak width consistency: Peaks that are significantly broader than expected (compared to the standard) may indicate on-column degradation, extra-column band broadening, or injection issues.
  4. Spectral purity (DAD/MS): If the purity index indicates spectral inhomogeneity across the peak, a co-eluting impurity is likely present. Consider adjusting the method for better resolution.
  5. Negative peaks or baseline disturbances: Negative peaks in UV indicate the sample solvent absorbs more than the mobile phase at the detection wavelength -- normal for the solvent front but abnormal elsewhere.
  6. Ghost peaks: Peaks present in the blank injection indicate carryover, contaminated mobile phase, or column bleed. Identify the source before reporting sample results.
  7. Summarize overall chromatographic quality and note any limitations on the reported results.

Got: Peak quality assessed for all target analytes; any anomalies (tailing, co-elution, ghost peaks) documented with their potential impact on data quality.

If fail: If significant quality issues are found (co-elution confirmed by spectral impurity, ghost peaks at analyte retention times), the data may not be reportable. Flag the results, investigate root cause, and re-run after corrective action.

Validation

Pitfalls

Related Skills